Agilent InfinityLab LC Series Vialsamplers User Manual
193
7
Error Information
General Error Messages
General Error Messages
Timeout
Error ID: 0062
The timeout threshold was exceeded.
Probable cause
Suggested actions
1
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
2
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.