Agilent InfinityLab LC Series Vialsamplers User Manual
184
6
Troubleshooting and Diagnostics
Maintenance and Troubleshooting Tools
Injector Steps
Injector Steps
is a diagnostic tool with which the user can walk through the whole
injection process while having manual control over the elemental steps. This can
be especially beneficial during troubleshooting when close observation of each
individual step of the injection cycle can be crucial for success.
The execution of the
Injector Steps
tool encompasses the following elementary
steps when the Agilent Lab Advisor Software is used:
1
Select
Injector Steps
from the list under
Service & Diagnostics
.
2
Enter a valid value for the vial location and the injection volume in the
Location
and
Volume
fields, respectively.
3
Select and execute any of the step commands (see
4
Select
Back
in the bottom right corner to exit the
Gripper Verification
function.
The vial is returned, the metering plunger moves into the home position, and
the injection valve switches into the mainpass position.
NOTE
For some commands, there are prerequisite steps that are executed
automatically if not already done (see