42
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
2
Site Requirements and Specifications
Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Considerations
The module power supply has wide ranging capability. It accepts any line
voltage in the range described in
on page 45. Consequently there is no
voltage selector in the rear of the module. There are also no externally
accessible fuses, because automatic electronic fuses are implemented in the
power supply.
WA R N I N G
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔
Connect your instrument to the specified line voltage only.
WA R N I N G
Electrical shock hazard
The module is partially energized when switched off, as long as the power cord is
plugged in.
The cover protects users from personal injuries, for example electrical shock.
➔
Do not open the cover.
➔
Do not operate the instrument and disconnect the power cable in case the cover
has any signs of damage.
➔
Contact Agilent for support and request an instrument repair service.
WA R N I N G
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the
power line at any time.
➔
Make sure the power connector of the instrument can be easily reached and
unplugged.
➔
Provide sufficient space behind the power socket of the instrument to unplug the
cable.