Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
Detector Error Messages
Wavelength Calibration Lost
Error ID: 6691
After exchanging the monochromator assemblies, the calibration factors
should be reset to defaults values (a new FLM board comes with default
values). In this case
“Wavelength Calibration Lost”
is displayed and the
instrument stays in a
Not Ready
condition.
Flow Cell Removed
Error ID: 6616, 6702, 6760
The detector has an automatic cell recognition system. When the flow cell is
removed, the lamp is turned off and a
NOT READY
condition exists. If the flow
cell is removed during an analysis, a
SHUT DOWN
is generated.
Probable cause
Suggested actions
1
Reset of monochromator settings after
exchange.
Perform a wavelength calibration.
2
Replacement of FLM board.
Perform a wavelength calibration.
Probable cause
Suggested actions
1
Flow cell has been removed during analysis.
Insert flow cell and turn on the lamp.