154
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
General Error Messages
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time,
the error message is generated.
Depending on the module, assemblies (e.g. the lamp in the detector) are
turned off to assure that the module does not overheat inside.
Open Cover
Error ID: 0205
The top foam has been removed.
Probable cause
Suggested actions
1
Fan cable disconnected.
Please contact your Agilent service
representative.
2
Defective fan.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Foam not activating the sensor.
Please contact your Agilent service
representative.
2
Defective sensor or main board.
Please contact your Agilent service
representative.