Optimizing the Detector
5
Method Development
Table 12
Conditions for simultanoeus multi wavelength detection for PNA-analysis (see
figure below)
Column
Vydac, 2.1 x 250 mm, PNA, 5 µm
Mobile phase
A = water; B = acetonitrile (50 : 50)
Gradient
3 min, 60 %
14.5 min, 90 %
22.5 min, 95 %
Flow rate
0.4 mL/min
Column temperature
22 °C
Injection volume
2 µL
FLD settings
PMT 12,
response time 4 s
The upper trace was
received with
traditional
wavelength
switching.
Figure 36
Simultaneous multi wavelength detection for PNA-analysis