![Agilent Technologies G7117C User Manual Download Page 86](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751086.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
86
5
Optimizing the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Table 17
Peak Width — Response Time — Data Rate (G7117B)
Peak width at half
height [min]
1
1
Values in the User Interface may be rounded.
Response
[s]
Signal data
rate [Hz]
Scan data rate
[HZ]
≤
126 pts/scan
Scan data rate
[HZ]
≤
251 pts/scan
Scan data rate
[HZ]
≤
501 pts/scan
Scan data rate
[HZ]
>501 pts/scan
< 0.00078125
0.0078125
240
240
80
40
20
> 0.00078125
0.015625
240
240
80
40
20
> 0.0015625
0.03125
160
160
80
40
20
> 0.003125
0.0625
80
80
80
40
20
> 0.00625
0.125
40
40
40
40
20
> 0.0125
0.25
20
20
20
20
20
> 0.025
0.5
10
10
10
10
10
> 0.05
1
5
5
5
5
5
> 0.1
2
2.5
2.5
2.5
2.5
2.5
> 0.2
4
1.25
1.25
1.25
1.25
1.25
> 0.4
8
0.625
0.625
0.625
0.625
0.625
> 0.85
16
0.3125
0.3125
0.3125
0.3125
0.3125
Table 18
Peak Width — Response Time — Data Rate (G7117A/G7117C)
Peak width at half
height [min]
1
Response [s]
Scan data
rate[Hz]
≤
251
pts/scan
Scan data
rate[Hz]
≤
501
pts/scan
Scan data
rate[Hz] > 501
pts/scan
< 0.0015625
0.015625
120
120
40
20
> 0.0015625
0.03125
120
120
40
20
> 0.003125
0.0625
80
80
40
20
> 0.00625
0.125
40
40
40
20
> 0.0125
0.25
20
20
20
20
> 0.025
0.5
10
10
10
10
> 0.05
1
5
5
5
5
> 0.1
2
2.5
2.5
2.5
2.5
> 0.2
4
1.25
1.25
1.25
1.25