![Agilent Technologies G7117C User Manual Download Page 171](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751171.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
171
9
Maintenance
Information from Module’s Assemblies
Information from Module’s Assemblies
Lamp and Flow Cell RFID Tag
The detector is equipped with a UV lamp and flow cell identification system using
RFID (radio frequency identification) tags attached to the assemblies and RFID
tag readers at the optical unit. The table below lists all parameters stored in the
RFID tag.
Serial Number and Firmware Revision
The user interface provides module specific information that is stored in the
main board. These are for example the serial number, firmware revision.
Table 26
RFID Tag Data
Lamp information
Flow cell information
•
product number
•
product number
•
serial number
•
serial number
•
production date
•
production date
•
accumulated UV on time (in hours)
•
nominal path length of the cell (in mm)
•
actual UV lamp on time (in hours)
•
cell volume (
) in
µ
L
•
number of ignitions
•
maximum pressure (in bar)
•
date of last intensity test
•
date of last cell test
NOTE
The pressure value is always displayed in bar, even if the user interface uses
other units, e.g. PSI.