![Agilent Technologies G7117C User Manual Download Page 159](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751159.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
159
9
Maintenance
Replace the Deuterium Lamp
Replace the Deuterium Lamp
When
If noise or drift exceeds application limits or lamp does not ignite.
Tools required
Description
Screwdriver POZI 1 PT3
Parts required
#
p/n
Description
1
5190-0917
Long-life Deuterium lamp (8-pin) with RFID tag
Preparations
Turn the lamp off.
WARNING
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
If so, wait for lamp to cool down.
NOTE
The lamp house cover includes a magnet.
1
Open the doors.
2
Locate the lamp cover.