Agilent InfinityLab LC Series Diode Array Detectors User Manual
127
8
Test Functions and Calibration
Conditions of Detector
Conditions of Detector
The test usually should be performed with a detector turned on for at least one
hour, so that the temperature regulation of the optical unit is working (not active
during the first 30 minutes after turn on). If the detector is on, tests can be
performed usually 10 minutes after the UV-lamp has been turned on.
Failing a Test
If a test fails with the Max-Light Cartridge Cell repeat the test with the Max-Light
Cartridge Test Cell and compare. If the test fails also, then start with proposed
actions mentioned in the details of the tests.