Agilent 1290 Infinity II Flexible Pump User Manual
6
Error Information
Pump Error Messages
Waste counter limit exceeded
Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been
set in the user interface.
•
Parameter: None
Flow rate limit exceeded
Error ID: 29164
The flow rate of the quaternary pump has exceeded the limit, while the
pump runs in pressure controlled mode, e.g. during a pressure test.
•
Parameter: None
Probable cause
Suggested actions
1
The waste container is full.
Empty waste container.
2
Inappropriate setting for waste counter.
•
Reset waste counter.
•
Adjust waste counter limit.
Probable cause
Suggested actions
1
Leak
Check for leaks in the pump and flow path.
2
Bottle empty.
Fill solvent bottle.
3
Shutoff valve closed (if applicable).
Open shutoff valve.
4
Drift of pressure sensor (unlikely for short
tests taking some minutes).
Replace pressure sensor.
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...