Agilent 1290 Infinity II Flexible Pump User Manual
141
Maintenance
7
Prepare the Pump Module for Transport
Prepare the Pump Module for Transport
When
If the module shall be transported or shipped.
Parts required
p/n
Description
9301-0411
Syringe; Plastic
9301-1337
Syringe adapter
G7104-44000
Transport protection foam
C A U T I O N
Mechanical damage
➔
For shipping the module, insert the Protective Foam to protected the module from
mechanical damage.
➔
Be careful not to damage tubing or capillary connections while inserting the module
in the Protective Foam.
1
Flush all solvent channels with isopropanol.
2
Remove solvent inlet tubes from solvent reservoirs and
tubing clips at other modules.
3
Remove tubings between the seal wash function and
solvent bottle/waste.
4
Turn off the pump.
5
Remove cable and capillary connections to other
modules.
6
Remove the waste tube.
7
Remove the module from the stack.
8
Disconnect the degasser outlet tubings at the MCGV one
after another. Use a syringe for removing liquid from the
degasser and the solvent tubings.
Summary of Contents for G7104A
Page 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Page 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Page 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Page 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Page 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Page 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...