Maintenance
Pump Head Procedures
Disassemble the Pump Head Assembly
When
If parts inside the pump head need to be replaced
Tools required
p/n
Description
G4220-20012
Torque wrench 2 – 25 Nm
G4220-20013
4 mm hex bit
G4220-20014
2.5 mm Hex Bit
G4220-20015
Adapter ¼ in square to hex
8710-0510
Wrench open 1/4 — 5/16 inch
Preparations
Remove the pump head
C A U T I O N
Damaged pump head
Disassembling or reassembling the pump head with tools other than the ones
recommended can damage pump heads and significantly reduce their life time.
➔
Follow all instructions step by step.
➔
Use recommended tools like the pump head alignment tool and a torque wrench.