Operating in Chemical Ionization (CI) Mode
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs
only
.
Materials needed
•
Octafluoronaphthalene (OFN), 100 fg/µL (5188-5347)
Procedure
1
Verify that the MSD performs correctly in EI mode.
2
Load the
NCICH4.U
tune file, and accept the temperature setpoints.
3
Select
Gas A
and set flow
to 40%.
4
In Tune and Vacuum Control view, run CI Autotune. See
NCI Autotune (Methane Reagent Gas)”
Note that there are no criteria for a “passing” Autotune in CI. If the
Autotune completes, it passes.
5
Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6
Verify that the system conforms to the published sensitivity specification.
Please see the Agilent Web site at
for
specifications.
C A U T I O N
Always verify MSD performance in EI before switching to CI operation. See
71. Always set up the CI MSD in PCI first, even if you are
going to run NCI.