110
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
To Switch from the CI Ion Source to the Extractor EI Ion Source
Procedure
1
From the
Tune and Vacuum Control
view, vent the MSD. See
76. The software will prompt you for the appropriate actions.
2
Open the analyzer. See
“To Open the Analyzer Chamber”
3
Remove the CI ion source. See
There is no need to remove the tip seal, as in the standard or inert EI ion
source. The CI interface tip seal will fit in the extractor EI ion source.
4
Install the EI extractor ion source. See
“To Install the EI Ion Source”
5
Locate the brown extractor wire from storage and connect it to the
extractor lens and the source board.
6
Place the CI ion source in the ion source storage box.
7
Pump down the MSD. See
“To Pump Down the MSD in EI Mode”
8
Load your EI tune file.
C A U T I O N
Always wear clean gloves while touching the analyzer or any other parts that go inside
the analyzer chamber.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap
and take other antistatic precautions
before
you open the analyzer chamber.
See
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......