Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
101
To Operate the Reagent Gas Flow Control Module
Procedure
1
In the
Manual Tune
dialog, click the
CI Gas
tab to access the parameter
settings for controlling the CI gas flow.
2
In the
Operations
area, select admit
a reagent gas
for the current tune file.
The system evacuates the gas lines for 6 minutes, then turns on the
selected gas (A or B). This is to reduce cross-mixing of the gases in the
lines.
C A U T I O N
After the system has been switched from EI to CI mode, or vented for any other reason,
the MS must be baked out for at least 2 hours before tuning.
C A U T I O N
Continuing with CI autotune if the MS has an air leak or large amounts of water will
result in
severe
ion source contamination. If this happens, you need to
vent the MS
and
clean the ion source
.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......