Starting up and shutting down
16
G5581A Microplate Labeler Quick Guide
After you finish loading new ribbon and labels
Use Microplate Labeler Diagnostics to run a print-and-apply test and verify that the
labels print correctly and the applicator head can pick up the labels from the printer.
Optionally, you can reset the label counter in the I/O Functions tab to track the number
of labels used.
Starting up and shutting down
Starting up the device
IMPORTANT
Follow the steps in the order shown to establish communication with the Microplate
Labeler.
To start up the Microplate Labeler
1
Turn on the air supply.
2
On the Microplate Labeler connection panel, check the air pressure gauge (see
figure,
1
). Verify the air pressure is set within the following range:
0.55 MPa (80 psi) to 0.58 MPa (85 psi)
Figure
Connection panel showing Ethernet connection
Note:
In Microplate Labeler Diagnostics, click the I/O tab to view the air pressure
readout after the pressure has been regulated by the internal pressure
regulator.
3
Verify that the communication cable (
2
) and power cord (
3
) are connected, but do
not turn on the power yet.
Note:
The figure shows an Ethernet connection. If you are using serial
communication, the serial cable connects to the serial port.
4
If you are using an Ethernet switch, turn on the power to the Ethernet switch.
1
2
3
4