1260 SFC Binary Pump
85
Error Information
6
Module Error Messages
Servo Restart Failed
Error ID: 2201, 2211
Servo Restart Failed:
Pump channel A
B: Servo Restart Failed:
Pump channel B
The pump motor in the module was unable to move into the correct position
for restarting.
When the module is switched on, the first step is to switch on the C phase of
the variable reluctance motor. The rotor should move to one of the C positions.
The C position is required for the servo to be able to take control of the phase
sequencing with the commutator. If the rotor is unable to move, or if the C
position cannot be reached, the error message is generated.
Probable cause
Suggested actions
1
Mechanical blockage of the module.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2
Disconnected or defective cable.
Please contact your Agilent service
representative.
3
Blocked (passive or active) inlet valve.
Exchange the (passive or active) inlet valve.
4
Defective pump drive assembly.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Summary of Contents for G4782A
Page 8: ...8 1260 SFC Binary Pump Contents...
Page 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Page 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Page 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Page 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Page 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...