Maintenance
Exchange the Solvent Selection Valve
Exchange the Solvent Selection Valve
When
If leaking internally (crossflow between the ports), or if one of the channels is blocked
Tools required
p/n
Description
8710-0899
Screwdriver, Pozidriv #1
Parts required
p/n
Description
G1381-60000
Solvent Selection Valve Upgrade Kit
Preparations
Switch off the pump at the main power switch
1
Lift solvent bottles out of the solvent cabinet and place
them on the table. Disconnect the solvent tubes from the
solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.
2
Disconnect all tubings from the solvent selection valve.
3
Using a Pozidriv screwdriver #1 loosen the holding
screws of the valve holder.
4
Carefully pull the valve holder out and disconnect the
valve cable at the connector.