1260 SFC Binary Pump
8
Maintenance
Exchange the Seal Wash Cartridge
Exchange the Seal Wash Cartridge
Parts required
p/n
Description
5065-4445
Peristaltic pump with Pharmed tubing
Preparations
•
Switch off pump at the main power switch.
•
Open the doors.
1
Remove the wash solvent tubings from the support ring
outlet and from the adapter leading to the waste bottle.
2
Unclip the peristaltic pump cartridge from the module
housing and remove it.