Advanced Functions
Check for Ready
Description
A hardwired connection is built into the remote start cable
between the GC and the Headspace Sampler. Use this function
to wait for the GC ready signal, abort the sequence, or proceed
without the GC ready signal.
If Check For Ready is active, then the sampler will check for the
presence of a ready signal from the GC just prior to injection. If
a GC ready signal is not found, the sampler waits, aborts, or
continues depending on the mode.
10. Reset
11. Valve Count
↓
ADV. FUNCT.
9. Check For Ready
No Check
5890 Wait
APG Wait
12. Vial Size
↑
↓
10. Reset
11. Valve Count
↓
ADV. FUNCT.
9. Check For Ready
APG Abort
5890 Abort
12. Vial Size
↑
↑