![Agilent Technologies G1888 User Information Download Page 102](http://html1.mh-extra.com/html/agilent-technologies/g1888/g1888_user-information_2867733102.webp)
102
User Information
Entering a Method
Storing and Loading Methods
Storing a method
Store Method places the method in active memory into one of
four storage locations for future use. To store the method
currently in use:
1
Press
Menu
. This displays the following screen:
2
Use the cursor keys to highlight
Store Method
.
3
Enter the desired storage location {1–4} on the numeric
keypad followed by
Enter
.
The Headspace Sampler stores the current method in the
specified memory location. A stored method includes the
following parameters:
•
Zone Temps
•
Event Times
•
Vial Parameters
•
The advanced functions for Oven Stabilization Time
•
The advanced functions for Pressure Units
•
The advanced functions for Multiple Headspace Extraction
DIAGNOSTIC
Store Method
Chain Methods
Pressures
↓
MAIN MENU
Load Method
READY
C A U T I O N
Storing overwrites any method already in that location. There is no
warning message that an overwrite is taking place.
medium_standard.book Page 102 Tuesday, February 17, 2004 10:14 AM