Optimizing Performance
2
1100 Series WS MWS Reference Manual
49
Fast Injection Cycle and Low Delay Volume
Short injection cycle times for high sample througput is one of the main issues
in analytical laboratories. Shortening cycle time starts with:
•
shortening column length
•
high flow rate
•
steep gradient
Having optimized these parameters, further reduction of cycle times can be
obtained using the overlapped injection mode.
Overlapped Injection Mode
In this process, as soon as the sample has reached the column, the injection
valve is switched back to bypass and the next injection cycle starts but waits
with switching to mainpass until the actual run is finished. You gain the
sample preparation time when using this process.
Switching the valve into the bypass position reduces the system delay volume
by approximately 300 µl (the mobile phase is directed to the column without
passing sample loop, needle and needle seat capillary). This can help to have
faster cycle times especially if low flow rates have to be used like it is
mandatory in narrow bore and micro bore HPLC.
N O T E
Having the valve in bypass position can increase the carry-over in the system.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......