Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
101
6. Needle into seat error
✔
Upgrade the firmware revision to A.04.14 or higher and the ChemStation
revision to A.08.04 or higher
✔
Check needle position and correct alignment in the pusher
✔
Perform an auto-alignment
•
Turn-off the system
•
Perform a forced cold start (move the 8-bit configuration dip switches 1,2,8
at the rear of the instrument to position 1)
•
Turn-on the system
•
Perform an auto-alignment
•
Turn-off the system
•
Move the 8-bit configuration dip switches 1,2,8 at the rear of the instrument
to position 0
•
Turn-on the system
✔
Check the connector from needle carrier to the sample transport unit
✔
Check the sample transport connectors on the main board/sample
transport
✔
Check the connector on the bottom of the sample transport unit
✔
Change the needle carrier assembly (G1367-60010)
✔
Change the sample transport unit (G1367-60019)
✔
Change the main board (G1367-69500)
N O T E
If the needle into seat error appears during the initialization of the Well Plate Sampler:
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......