Agilent 1100 Series Fraction Collectors User’s Guide
97
Specifications
5
Performance Specifications for the Fraction Collectors
Delay volumes [
µ
l]
Fraction collector inlet to diverter valve: ~500 (typical, depends on length of
the tubing)
Diverter valve: ~15
Diverter valve to needle: ~110
Needle: ~5
Delay calibration
sensor
Single wavelength absorbance detector working at 654 nm, consisting of a
LED and a photo diode
Diverter valve
3/2 Diverter valve with low internal volume (15
µ
l), switching time
< 100 ms, maximum operating pressure 6 bar
cooling
Optional (with additional G1330B), performance depending on ambient
conditions and the volume of collected fractions
maximum capacity
3 fraction collectors in parallel plus one recovery fraction collector
connected via 12-Position, 13-Port Selector valve (PN G1160A)
GLP features
Early maintenance feedback (EMF), electronic records of maintenance and
errors
Interfaces
Controller-area network (CAN).
optional; LAN or external contacts interface
RS232C,
APG-remote (for remote start / stop signals to / from other modules)
Interface to G1330A Thermostat
CAN-DC-out for operation of Agilent
approved external devices like valves
Safety features
Leak detection and safe leak handling, error detection and display, exhaust
fan for fume extraction of hazardous vapors
* Vials can be used as recommended by Agilent Technologies (see
“List of Recommended Vials and
“List of Recommended Plates and Closing Mats”
but must be
uncapped
. Only the 96 deep well-plates can be used (
without closing mats
, see
mended Plates and Closing Mats”
Table 23
performance Specifications Agilent 1100 Series PREPARATIVE SCALE Fraction
Collector (G1364B) (continued)
Type
Specification
N O T E
Only one type of well-plates can be used at a time in one tray.