Installing the Autosampler
3
Installing the Thermostatted Autosampler
1
Place the Thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste.
Figure 15
Condensation Leak outlet
3
Remove the adhesive tape which covers the front door.
4
Remove the front door and remove the transport protection foam.
5
Place the Autosampler on top of the Thermostat. Make sure that the
Autosampler is correctly engaged in the Thermostat locks.
WA R N I N G
Personal injury
To avoid personal injury, keep fingers away from the needle area during
Autosampler operation.
➔
Do not attempt to insert or remove a vial or a plate when the needle is positioned.
WA R N I N G
Damage through condensation
If the condensation tube is located in liquid the condensed water cannot flow out of
the tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
➔
Make sure that the condensation tube is always above the liquid level in the vessel.
8dcYZchVi^dcYgV^cijWZ
LVhiZWdiiaZ
Summary of Contents for G1329A
Page 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Page 6: ...6 Contents...
Page 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Page 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Page 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Page 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Page 136: ...136 9 Identifying Cables RS 232 Cables...
Page 149: ...149 Index...