Using the Manual Injection Valve
About the Injection Valve
About the Injection Valve
The injection valve has six ports and can be moved manually between two
positions. In the LOAD position, the mobile phase is led from the pump
directly to the column, allowing you to inject your sample into the sample loop.
In the INJECT position, the mobile phase is led through the sample loop,
flushing the contents of the loop onto the column.
Valve Positions:
Figure 15
Valve Positions
1 to 6
valve ports
7
inlet (from pump)
8
outlet (to column)
9
needle port
10
vents
11
sample loop