1200 Series Handheld Control Module Users’s Guide
139
Running Multiple-Vial Analyses
11
Multiple-Level Calibration Sequences
Multiple-Level Calibration Sequences
The following sections describe how to set up calibration sequences for
analyses which use multiple-level calibration.
Recalibrating With the Same Group of Standards
There are three calibration standards of different concentrations (C1, C2, C3)
and 15 samples (S). The standards and samples are analyzed using the same
method.
The analysis requires that:
•
each sample is analyzed once,
•
the calibration standards are analyzed twice before the samples and
re-analyzed twice after every 5 samples,
C1 C2 C3 S10-S14 C1 C2 C3 S15-S19 C1 C2 C3 S20-S24 C1 C2 C3
•
the calibration standards are located in positions 90 to 92 of the
autosampler tray,
•
the 15 sample vials are located in positions 10 to 24 of the autosampler tray,
and
•
the samples and standard are analyzed using a method called METH1.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the
Delete
button [
F6
].
1
Select the
Sequence
button [
F4
] in the
Analysis
screen.
2
For sequence line 1, select the
Insert
button [
F7
] and
enter
:
Vial Range
10 to 24
#Inj.
2
Inj. Volume
Default
Method
METH1