Agilent 1260 Infinity VWD User Manual
61
Using the Detector
4
Setting up an Analysis
Figure 19
Detector Settings (default)
•
1 signal with individual wavelength setting
•
stop and post time can be set (if required)
•
peakwidth depends on the peaks in the
chromatogram, see
•
time table for programmable actions during the run
•
Zero Offset Limits: 1 – 99 % in steps of 1 %
•
Attenuation Limits: 0.98 – 4000 mAU at discrete
values for either 100 mV or 1 V full scale
•
additional signals can be stored with the normal
signal (for diagnostics)
•
autobalance to zero absorbance (on the analog
output plus offset) at begin and/or end of run
•
see