Agilent 1260 Infinity VWD User Manual
25
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications
Performance Specifications G1314B
Table 3
Performance Specifications G1314B
Type
Specification
Comments
Detection type
Double-beam photometer
Light source
Deuterium lamp
Wavelength range
190 – 600 nm
Short term noise (ASTM)
<± 0.5·10
-5
AU at 254 nm
Drift
3·10
-4
AU/h at 254 nm
Linearity
> 2 AU (5 %) upper limit
See
Wavelength accuracy
± 1 nm
Self-calibration with
deuterium lines, verification
with holmium oxide filter
Maximum data rate
13 Hz
Band width
6.5 nm typical
Flow cells
Standard: 14 µL volume,
10 mm cell path length and
40 bar (580 psi) pressure
maximum
High pressure: 14 µL volume,
10 mm cell path length and
400 bar (5800 psi) pressure
maximum
Micro: 1 µL volume, 5 mm cell
path length and 40 bar
(580 psi) pressure maximum
Semi-micro: 5 µL volume,
6 mm cell path length and
40 bar (580 psi) pressure
maximum
Can be repaired on component
level