Agilent 1260 Infinity VWD User Manual
167
Hardware Information
12
Firmware Description
•
Instant Pilot (G4208A) with files from a USB Flash Disk
•
Agilent Lab Advisor software B.01.03 and above
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315AB for the G1315A/B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C
DAD,
VVV is the revision number, for example 102 is revision 1.02,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section
Replacing Firmware
in chapter
"Maintenance"
or use the documentation provided with the
Firmware Update Tools
.
Figure 51
Firmware Update Mechanism
N O T E
Update of main system can be done in the resident system only. Update of the resident
system can be done in the main system only.
Main and resident firmware must be from the same set.
GZh^YZciHnhiZb
BV^c;LjeYViZ
BV^cHnhiZb
GZh^YZci;LJeYViZ