84
Troubleshooting and Test Functions
Park Arm (Park Gripper)
Park Arm (Park Gripper)
C A U T I O N
Before transporting or shipping the autosampler, always secure the arm in the
park position.
In order to prevent mechanical damage to the transport mechanism during
transport, the gripper arm should be moved into the park position. The park
arm function moves the gripper and transport slider to the home position
behind the sampling unit, and lowers the gripper arm into the park position
where the transport assembly is secured against a mechanical stop. The
autosampler can be switched off after parking the arm.
N O T E
Before parking the gripper arm, ensure there is no vial in the gripper. Use the
“Release Gripper” function to remove the vial.
User Interface
In the ChemStation the Park Arm command is part of the ALS maintenance
positions that can be selected from the Maintenance menu in the Diagnosis
display. In the Control Module the Park Gripper command is located in the
Control display of the autosampler.
The commands for the ChemStation and Control Module (different names for
commands in the Control Module are shown in brackets) are:
Park Arm
(Park Gripper)
moves the gripper arm to the park position.
Home
moves the gripper arm out of the park position to the
home position.
Using the Park Arm Function
1
Select “Park Arm” (“Park Gripper”).
2
When the arm is in the park position, the autosampler is ready for shipment,
and can be switched off.
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......