55
Troubleshooting and Test Functions
Shutdown
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The autosampler continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
•
Leak detected in an external instrument with a remote connection to the
system.
•
Shut-down in an external instrument with a remote connection to the
system.
•
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
❏
Fix the leak in the external instrument before restarting the autosampler.
❏
Check external instruments for a shut-down condition.
❏
Check the degasser for an error condition. Refer to the
Reference Manual
for the Agilent 1100 Series degasser.
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......