Repairing the Autosampler
Metering-Drive Motor and Belt
Metering-Drive Motor and Belt
Frequency
When belt or motor defective
Tools required
Pozidrive No. 1 screwdriver
Parts required
Metering-drive motor 5062-8590
Belt 1500-0697
Before beginning this procedure:
❏
Remove the top cover (page 128).
❏
Remove the transport assembly (page 132).
❏
Remove the sampling unit (page 134).
7
Disconnect the motor connector from the
SUD board (labelled “Metering M”).
8
Remove the four fixing screws.
9
Remove the motor.