Flow Tracker Series 1000/2000
Reference Manual
23
Leak mode (Flow Tracker 2000 only)
The Agilent Flow Tracker 2000 flowmeters use the thermal
conductivity property of gases to detect leaks. The mechanism
for detecting leaks is entirely separate from the main flow
metering portion of the device. It consists of a sample suction
probe, internal flow channel with thermal conductivity sensors,
an internal suction pump, and an exhaust port.
Figure 4
Leak detector mode screen/buttons
GAS
MODE
OFF
ON
ZERO
Gas select
Power off
Mode
Mode indicator
Low battery indicator
Sound on/off
Leak rate
Thermal conductivity
Relative conductivity
Primary display
(default is leak rate)
Power on
Zero (tare)