Flow Tracker Series 1000/2000
Reference Manual
19
Gas select
The selected gas is shown in the upper right hand corner of the
display. There are seven different gases commonly used for gas
chromatography that can be selected. The gas can be changed by
pushing the
GAS
button until the desired gas is displayed. The
selected gas is common to all operating modes and can be
changed at any time in any operating mode. The gases are
shown in
Low battery indicator (LO BAT)
The low battery indicator, located just above the mode indicator
in the lower right corner of the display, appears if the batteries
run below a certain voltage level that can affect the accuracy of
the flowmeter.
Flow mode indicator
This indicator, located in the lower right corner of the display,
signals that the unit is in Flow mode. The operating mode can
be changed by pushing the
MODE
button.
0
°
C
and
14.695 PSIA
25
°
C
and
14.695 PSIA
0
°
C
and
760 torr (mmHG)
70
°
F
and
14.695 PSIA
68
°
F
and
29.92 inHG
20
°
C
and
760 torr (mmHG)
N O T E
The Agilent Flow Tracker Series flowmeters reference 25 ºC and
14.695 PSIA (101.32 kPa).
C A U T I O N
To avoid inaccurate readings, be sure to change the batteries when the
LO BAT indicator is displayed. Low power can result in inflated
temperature sensor readings which affect the expected gas viscosity
and mass flow calculations.