Power Meter (USB) Mode
147
Averaging
Averaging is used to improve measurement accuracy in low power or noisy
power situations. Increased averaging improves measurement accuracy, but also
reduces measurement speed.
How to set Power Meter Averaging
Press
BW 2
or
Meas Setup 4
.
Then choose from the following:
o
Averaging Auto Man Off
o
Auto
Allows the USB Power Sensor to use its Auto Averaging settings.
See the USB Power Sensor documentation for details.
o
Man
Manually enter an averaging count by setting
Num Averages
.
o
OFF
Performs NO averaging.
o
Num Averages
Sets the number of Power Meter readings to be averaged
before a valid reading is displayed. When a number is entered,
Averaging
is
automatically set to Man.
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
Press
Sweep 3
.
Then choose one of the following:
o
Single
Automatically sets Continuous OFF and causes FieldFox to make
ONE measurement, then hold for the next Single key press. When a data
trace is displayed, the entire trace is measured, then holds. The
Hold
annotation changes to an arrow
-->
while the measurement occurs.
o
Continuous
Makes continuous measurements. This is the typical setting
when battery power is not critical.
You can also use
Run / Hold +/-
to toggle between Single and Continuous.
Step Detection Mode
The FieldFox supports the Step Detection feature that is present in Agilent USB
power sensors. When enabled, this feature reduces settling time after a
significant step in the measured power. For more information, please refer to the
USB Power Meter documentation.
How to set Step Detection mode
Press
Meas Setup 4
.
Then
Step Detection ON Off