SA (Spectrum Analyzer) Mode
97
When you
first select a Channel Measurement
, then select
a Radio Standard
,
the BW, Offset, RRC, Integration BW, and Span settings are changed to those of
the standard. In addition, Res BW can also change when set to ‘Auto(couple)’.
However, center frequency is NOT changed unless you first select Units = CHAN.
Measurement Preset
Measurement Preset allows you to easily reset any of the channel measurements
to its default settings. The Center Frequency, Preamp ON|OFF, RF Attenuation,
Markers, Limits, and Radio Standard settings are NOT reset.
To select Measurement Preset:
Press
Preset
then
Meas Preset
Averaging
By default in ALL Channel measurements, averaging is enabled and set to
display the average of the last 15 measurements. When enabled, this average
setting is automatically making the following ‘averaging’ settings in order to
provide the most accurate power measurements:
Averaging Type is set to Power.
Trace State is set to Average.
Detector is set to Average.
Count is set to 15.
Any of these settings can be changed manually during a Channel measurement.
Learn more about these settings starting at page 90.
To change Averaging:
Press
Meas Setup 4
.
Then
Average Enable ON OFF
OFF effectively sets the Average Count to 1.
Then
Average Count
Enter a value between 1 and 10,000 using the numeric keypad,
▲|▼
arrows,
or the rotary knob.
Press
Enter
.
Traces
Only one measurement trace can be displayed in Channel Measurements.
Channel Power (CHP)
Channel Power measures total power over the specified Integrated BW. The
Integration Bandwidth (IBW) can be adjusted to measure the power over
multiple channels.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...