N9912A FieldFox User’s Guide
Then choose from:
o
M (meters)
o
Feet
Calculated DTF values
Press
Next Page
on the DTF Settings Table to view the following
calculated
Values
noted on the FieldFox screen with c - <setting>
:
Start Frequency
– Start frequency that is used to calculate DTF.
Stop Frequency
– Stop frequency that is used to calculate DTF.
Range Resolution
. Indicates the accuracy of the distance to fault measurement.
For example, with range resolution of 500 mm, if the distance to fault is 10
meters, this value could be inaccurate by +/- 500 mm or between 9.5 to 10.5
meters. This value is calculated from frequency span / resolution (points).
Response Resolution, not displayed, indicates the distance that could be between
two faults and still show as separate faults. Learn more in Window Settings on
page 37.
Maximum Distance
. The distance that could be viewed with the current settings.
Defined by: Vf*c*Points/(2*Bandwidth) where:
o
Vf = velocity factor
o
c = speed of light
o
Points = resolution
o
Bandwidth = frequency range
Alias-free Range
(On/Off)
o
On = No Alias images
o
Off = Alias images may appear in the response.
About Alias Faults
An alias fault is not a true device response. An alias fault appears because of the
method used to convert frequency to time.
On the DTF Settings page (above) the
c - Alias-free Range = Off
setting indicates
alias images MAY appear on the screen.
Shorter stop distances (less than 10 meters) and a higher resolution (1001
points) will be
more
likely to result in
Alias-free Range = Off
.
When the Alias-free Range = Off, the following procedure will help to determine
if a response is true or an alias response:
1. Put a marker on the response in question and note the distance to the fault.
2. Change the start or stop distance.
A true fault response will not move in distance. That is, if a true fault is present
at 10.3 meters, changing the stop distance from 15 m to 20 m will not move the
fault; the fault will remain at 10.3 meters. However, an alias response will
appear to move.
Summary of Contents for FieldFox N9912A
Page 15: ...Preparing for Initial Use of Your New FieldFox 15 Take the FieldFox Tour Front Panel ...
Page 194: ...194 N9912A FieldFox User s Guide Batteries Safe Handling and Disposal ...
Page 195: ...Safety Considerations 195 Inspired Energy Battery ...
Page 196: ...196 N9912A FieldFox User s Guide ...
Page 197: ...Safety Considerations 197 ...
Page 198: ...198 N9912A FieldFox User s Guide ...