N9927-90003
7-35
FieldFox Analyzer
Repair and Replacement Procedures
Removing and Replacing the Measurement Group Assemblies
Replacing the A5 System Board
1.
Separate the front and rear cases. See
“Separating the Front and Rear Cases” on page 7-10
.
2. Remove all measurement group assemblies as one unit. Refer to
Group Assemblies as One Unit” on page 7-27
3. Remove the I/O side panel. Refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page
4. Disconnect cable W3 and W4 from the A5 System board.
5. Refer to
. Remove the RTC board. Refer to
“Replacing the Real Time Clock (RTC)
Battery Located on the A5 System Board Shield” on page 7-31
6. Remove the System board shield by removing the nineteen screws (item 2). Discard the screws, new
ones are provided in the
7. Refer to
. Remove the four screws (item 2) that secure the A6 SOM board (item 1) to the
A5 System board. Discard these screws; new ones are provided in the
8. Unplug the A6 SOM board from the A5 System board.
9. Refer to
. Remove the two screws (item 1) from the A5 System board, lift it up from the
RF bottom shield, and unplug it from the A4 RF board connector. Discard the screws, new ones are
provided in the
.
10. Refer to
. Plug in the new A5 System board to the A4 RF board, over the new gasket, and
aligned with the three dowel pins in the RF bottom shield. Make sure the bullets are straight up when
connecting the System and RF boards.
. Replace the SOM board onto the System board. Install four new screws (item 2)
from the
and tighten them in the torque sequence indicated by the circled
numbers (1 - 2 - 3 - 4) in the illustration.
12. Refer to
. Place the System shield onto the System board by aligning the dowel pins.
Replace the nineteen shield screws. Torque these screws (plus the two System board screws, #13 and
#21) in the sequence shown by the circled numbers (1 - 21). Repeat this torque sequence one
additional time. Torque to 12 in-lbs.
13. Place the RTC board onto the System board shield and secure with the two screws. Torque to 9 in-lbs.
14. Remove the System board to FPIB cable (W14) from the old A5 System board and install it on the
new board.
15. Reinstall the I/O side panel. Refer to
“Replacing the I/O Side Panel, Doors, and Gasket” on page
16. Perform the steps under
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...