7-30
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Removing and Replacing the Measurement Group Assemblies
Replacing the Type-N RF Connectors and RF Cables
1.
Separate the front and rear cases. See
“Separating the Front and Rear Cases” on page 7-10
2. Remove all measurement group assemblies as one unit. Refer to
Group Assemblies as One Unit” on page 7-27
.
3. Remove the RF top assembly. Refer to
“Replacing the RF Top Panel Assembly” on page 7-29
.
4. Refer to
. Disconnect the cable connector nut on cable W1 or W2, depending on which
connector or cable is being replaced. Do not attempt to remove the cable, just release the connector
nut.
5. Remove the two screws (item 2) from the type-N RF connector to be replaced.
6. Lift the type-N RF connector upwards and out of the RF bottom shield.
7. The RF input cable can now be replaced by disconnecting the other end of the cable and removing it.
When installing the new cable, do not tighten either of the cable connector nuts until both of them
have been installed finger tight.
8. Remove the RFI O-ring.
9. Remove the old insulator and install a new one.
10. Position the new type-N RF connector in the RF bottom shield, making sure that it aligns with the
cable connector underneath. Loosely install the cable connector nut.
11. Install a new RFI O-ring over the new type-N RF connector in the position shown.
12. Loosely reinstall the two screws (item 2). Note the location for these two screws in the illustration.
To ensure proper alignment, install two additional screws, finger tight, in the two empty holes in the
type-N RF connector, then tighten the two screws that secure the connector. Remove the two
alignment screws.
13. Tighten the cable connector nut(s) on cable W1 or W2.
14. Perform the steps under
“Post-replacement Procedure” on page 7-44
.
Figure 7-20
Replacing the Type-N RF Connectors and RF Cables
Summary of Contents for FieldFox MW N9913A
Page 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Page 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Page 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Page 17: ...N9927 90003 2 1 2 General Product Information ...
Page 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Page 37: ...N9927 90003 4 1 4 Troubleshooting ...
Page 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Page 51: ...N9927 90003 5 1 5 Theory of Operation ...
Page 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Page 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Page 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Page 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Page 149: ...N9927 90003 Appendix 1 Appendix A ...
Page 156: ...Index 4 N9927 90003 Index ...