7-10
N9927-90003
Repair and Replacement Procedures
FieldFox Analyzer
Removing and Replacing the Case Group Parts
Separating the Front and Rear Cases
1. Remove the battery. Refer to
“Removing and Replacing the Main Battery” on page 7-6
2. Refer to
. Loosen the six screws (item 3) that secure the RF top panel. It should not be
necessary to remove these screws, just loosen them.
3. Remove, and discard, the six screws (item 4) that secure the rear case to the front case. New screws
.
Figure 7-4
Separating the Front and Rear Cases (1)