Eyesys Mini-B/A Gauge Controller
18
Remote Reset
A remote reset can take place whether in Local or Remote control. As with a power-up
reset, remote reset resets the following parameters:
Remote reset however, leaves the Remote/Local Control status untouched, whereas a
power-up reset will reset to Local Control.
Command/Response Format
Command Set
All lower-case characters are place holders and must be replaced as follows:
Gas Correction to 1.000
Sensitivity to 15.0
The command format is:
“#” {Gauge address} {command} {data} {carriage return}
The response is:
“>” {optional data} {carriage return}
aa
= two character listener address (00 for RS232)
x
= “1”..“9” data value
{cr}
= message terminator character (carriage return)
Table A-1
Command Set
Command
Response
Read AG Type
#aa01{cr}
>21FEFEFEFE{cr}
Read Pressure
#aa02I1{cr}
>x.xxxE-xx{cr}
Read Setpoint State
#aa03{cr}
>000x{cr}
where
x
Relay1
Relay2
0
OFF
OFF
1
ON
OFF
2
OFF
ON
3
ON
ON