Real-Time I/Q Baseband Generator User’s and Programming Guide
6-9
ESG Family Signal Generators
Theory of Operation
Options UN8 and 202
Digital Modulation Input/Output Relationships
Digital Modulation Input/Output Relationships
NOTE
With Option 1EM, all connectors are located on the rear panel.
The baseband generator’s clock can be internally or externally supplied, and the external
data clock can be set to a normal bit clock or a symbol clock. Combinations of these
selections will affect the inputs required and the outputs available as shown in the
following table.
Table 6-3. Data Pattern Mode
Front Panel Settings
Front Panel Inputs
Rear Panel Outputs
Data
BBG
Data
CLK
Ext
Data
Clock
DATA
CLK
SYMB
SYNC
DATA
DATA
CLK
OUT
SYMB
SYNC
OUT
DATA
OUT
Int
Int
NA
NA
NA
NA
Internal
Bit Clock
Internal
Symbol
Clock
Internally
Generated
Data
Int
Ext
Normal
User’s
Bit
Clock
NA
NA
Internal
Bit Clock
Internal
Symbol
Clock
Internally
Generated
Data
Int
Ext
Symbol
User’s
Symbol
Clock
NA
NA
Internal
Bit Clock
Internal
Symbol
Clock
Internally
Generated
Data
Ext
Int
NA
NA
User’s
Symbol
Clock
a
a. The front panel SYMBOL SYNC and DATA inputs must be clocked by the
internal bit clock output from the rear panel DATA CLK OUT.
User’s
Ext
Data
Internal
Bit Clock
User’s
Symbol
Clock
User’s
External
Data
Ext
Ext
Normal
User’s
Bit
Clock
User’s
Symbol
Clock
User’s
Ext
Data
Internal
Bit Clock
User’s
Symbol
Clock
User’s
External
Data
Ext
Ext
Symbol
User’s
Symbol
Clock
b
b. When the user’s symbol clock is supplied to the DATA CLOCK input, it must also
be supplied to the SYMBOL SYNC input by means of a tee.
User’s
Symbol
Clock
User’s
Ext
Data
Internal
Bit Clock
User’s
Symbol
Clock
User’s
External
Data