4-72
Real-Time I/Q Baseband Generator User’s and Programming Guide
Programming Commands and Examples
ESG Family Signal Generators
Programming Commands
Options UN8 and 202
synchronized to the bit selected by the synchronization output offset of a
framed data pattern.
SLOT
Outputs a 1-bit signal to the EVENT 1 rear panel connector that is
synchronized to the bit selected by the synchronization output offset of a
selected timeslot.
ALL
Outputs a 1-bit signal to the EVENT 1 rear panel connector that is
synchronized to the bit selected by the synchronization output offset of
each timeslot in a frame.
*RST value: Frame
Synchronization Output Offset
[:SOURce]:RADio:<desired format>:SOUT:OFFSet <value>
[:SOURce]:RADio:<desired format>:SOUT:OFFSet?
This command moves the synchronization signal forward or back from the beginning of the
data frame or timeslot.
Range:
• DECT, EDGE, GSM:
−
155 t155 bits
• NADC:
−
323 t323 bits
• PDC:
−
279 t279 bits
• PHS:
−
239 t239 bits
• TETRA:
−
509 t509 bits
*RST value: 0 bits