4-12
Real-Time I/Q Baseband Generator User’s and Programming Guide
Programming Commands and Examples
ESG Family Signal Generators
Programming Commands
Options UN8 and 202
Burst Rise Time
[:SOURce]:RADio:CUSTom:BURSt:SHAPe:RISE:TIME <value>
[:SOURce]:RADio:CUSTom:BURSt:SHAPe:RISE:TIME?
[:SOURce]:RADio:CUSTom:BURSt:SHAPe:RTIMe <value>
[:SOURce]:RADio:CUSTom:BURSt:SHAPe:RTIMe?
[:SOURce]:RADio:<desired format>:BURSt:SHAPe:RISE:TIME <value>
[:SOURce]:RADio:<desired format>:BURSt:SHAPe:RISE:TIME?
[:SOURce]:RADio:<desired format>:BURSt:SHAPe:RTIMe <value>
[:SOURce]:RADio:<desired format>:BURSt:SHAPe:RTIMe?
This command sets the burst shape rise time.
Range: minimum and maximum values depend upon modulation type and symbol rate.
*RST value:
• Custom: 5.860
• DECT: 13.797 bits
• EDGE: 12.700 bits
• GSM: 4.224 bits
• NADC: 5.860 bits
• PDC: 5.470 bits
• PHS: 4.976 bits
• TETRA: 9.376 bits
PN9 Mode
[:SOURce]:RADio:<desired format>
1
:BURSt:PN9 NORMal|QUICk
[:SOURce]:RADio:<desired format>
1
:BURSt:PN9?
1
All formats except EDGE.
This command controls the software PN9 generation. Normal mode produces a maximal
length PN9 sequence. Quick mode produces a truncated PN9 sequence. Use Normal mode
for bit-error-rate tests where a maximal length PN9 sequence is required.
*RST value: Normal
Cell Station Identification Code (CSID) Field
[SOURce]:RADio:PHS:DLINK:SLOT[1]|2|3|4:SCHannel:CSID <bit pattern>
[SOURce]:RADio:PHS:DLINK:SLOT[1]|2|3|4:SCHannel:CSID?
[SOURce]:RADio:PHS:ULINK:SLOT[1]|2|3|4:SCHannel:CSID <bit pattern>
[SOURce]:RADio:PHS:ULINK:SLOT[1]|2|3|4:SCHannel:CSID?
This command enables you to change the 42-bit cell station identification code (CSID) field.
The preset hexadecimal value (when normal preset is selected) for CSID reflects the PHS
protocol, however you can enter a new value with this command. The current value for
CSID is displayed in the
CSID
field near the bottom of the text area of the display.
Range: 00000000000 to 3FFFFFFFFFF (hexadecimal)
*RST value: 20200020001