Real-Time I/Q Baseband Generator User’s and Programming Guide
3-155
ESG Family Signal Generators
Softkey Reference
Options UN8 and 202
S
SW
Press this softkey to make the 20-bit frame synchronization word the
active function. The preset hexadecimal value (when normal preset is
selected) for SW reflects the PDC format, and appears in the
SW
field
near the bottom of the text area of the display.
Default Values: 785B4 (Up TCH and Up VOX), 87A4B (Dn TCH)
Softkey Location:
Mode
>
TDMA
>
PDC
>
Data Format Pattern Framed
>
Configure Timeslots
>
Configure Up TCH
>
SW
SW
is also located in the Configure Down TCH and Configure Up
VOX menus.
Symbol Rate
Press this softkey to make the transmission symbol rate the active
function.
Range
EDGE: 50.863 sps through 8.333333333 ksps
Other Formats: 47.684 sps through 12.500000 Msps
NOTE
The range varies, depending on the modulation type and the data that
is being transmitted.
Default Values
Custom, NADC: 24.300 ksps
PDC: 21.000 ksps
DECT: 1.152000 Msps
PHS: 192.000 ksps
EDGE and GSM: 270.833 ksps
TETRA: 18.000 ksps
Softkey Locations
Custom:
Mode
>
Custom
>
Symbol Rate
>
Symbol Rate
Other Formats:
Mode
>
TDMA
>
desired format
>
Modify Standard
>
Symbol Rate
>
Symbol Rate
Symbol Sync Out
Press this key to access a menu from which you can modify the output
at the SYMBOL SYNC OUT rear panel BNC/SMB connector.
Softkey Location:
Mode
>
TDMA
>
EDGE (Rev 8.3.0Release 1999)
>
BNC MUX Settings
>
Symbol Sync Out