13116
Search Length is clipped to the upper value according to the
dependency of IF Band Width
Search length upper limits:
IF Band Width <= 1 MHz, Limit = 0.1 sec
1MHz < IF Band Width <= 10MHz, Limit = 0.05 sec
10MHz < IF Band Width <= 20MHz, Limit = 0.025 sec
20MHz < IF Band Width <= 40MHz, Limit = 0.013 sec
40MHz < IF Band Width, Limit = 0.007sec
13118
Search Length is clipped to the lower value as allowed by
current IF Band Width.
Search length lower limits:
IF Band Width <= 10MHz, Limit = 0.001 sec;
10MHz < IF Band Width Limit = 0.1
µ
sec
13120
Symbol Rate is clipped to its limit because of IF Path change.
Upper limit of Symbol Rate is 10 MHz when IF Path is Narrow.
13122
IF BW is clipped to its limit because of IF Path change.
Limit of IF BW depends on IF Path selection. If the value exceeds the
limit, it is clipped to the limit. This message will not appear when IF BW is
set to AUTO.
13124
Meas Interval is clipped to its upper limit because of
points/symbol change.
Upper limit of Meas Interval depends on points/symbol setting.
13126
Only one marker is available on Demod Bits trace.
Only one marker is available on Demod Bits trace.
13201 to 13300: Stimulus-Response Measurements Error Messages (Option 215)
An error number in this range indicates the instrument has detected an error relating to
the Stimulus Response measurement personality.
13226
Connection could not be established with specified IP Address.
A communication connection could not be established at the specified IP
Address. Check that the IP address entered is correct and that the signal
generator and PSA are both connected to the LAN or are connected to one
another using a LAN cross-over cable. If you are using a LAN cross-over
cable the IP address of both instruments must start with the same first
two integers. For example, if the analyzer IP address is 156.121.102.33
then the source IP address must be 156.121.xxx.xxx.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...