10192
Sweep Point/Span ratio below minimum - results may be
inaccurate.
The sweep point to span ratio is below the minimum required to ensure
the bucket ratio is large enough to test DVB-T masks
10218
Hardkeys are disabled.
Some of the forms (for example Receiver Spurious in GSM) do not allow
you to close the form without either formally accepting or cancelling the
form settings. For this reason, all of the hardkeys are disabled until you
terminate the form.
10219
Awaiting trigger
The measurement requires a trigger to be present. If the trigger does not
occur or is delayed, this message will be displayed. Check your trigger
settings.
10220
Preselector cannot be bypassed in this mode.
The Option 123 allows the microwave preselector to be bypassed. This
functionality is not available from within either the Phase Noise or Noise
Figure application modes. If the preselector is set to off, it will
immediately be forced back to On.
10227
Measurement suspended until carrier is turned off.
The receive channel power and the receive spur measurements are
specified with the attenuation set to 0 dB. To prevent overload, the
frequency spectrum of interest is monitored for signal levels which exceed
a specified threshold before setting the attenuator to 0 dB. If a carrier is
found, this message is displayed and the completion of the measurement
will not occur until the carrier is removed. The carrier check may be
turned off using the properties form under the front-panel
Mode Setup
key.
You may also change the signal threshold which determines a carrier on
the properties form.
10228
Cannot correlate to input signal
.
This error is normally generated because of one of the following reasons:
1. There is no carrier signal.
2. Walsh channels other than the pilot are active.
3. There is some other modulation problem that will prevent the
measurement from being made.
This problem must be corrected before the measurement can continue.
10229
The regression portion failed.
This message occurs when (Option B7D) is not functioning properly.
Demodulation measurements (modulation accuracy and code domain)
might fail as a result of this error.
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...