54
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
603 PSA only
615 ESA only
File is corrupt
Corrupted file
The file that you were trying to load is corrupt.
617
Wrong density floppy inserted
The floppy disk has the wrong density. It should be 1.44 MB.
618
Illegal write access of Flash memory
Attempt to write to an unavailable area of internal flash memory.
619
Can’t Auto-Couple RBW in Zero Span
You sent a remote command to set the RBW into auto while in zero span.
620
Can’t Auto-Couple Sweep Time in Zero Span
You sent a remote command to set the sweep time to auto while in zero
span.
622
External reference missing or out of range
The external frequency reference signal is missing, has too low an
amplitude, or does not match the frequency value that you previously
entered into instrument memory.
623
Printer not responding
Printer not responding; ioctl ERROR
Check the printer. It may not be connected properly or turned on.
625
Printer out of paper
Load paper in printer.
626
Print failed; Can’t open print file desc
Print failed; DisplayExport can’t create print file
Print failed; DisplayExport illegal print language
Print failed; DisplayExport insufficient RAM space
Print failed; DisplayExport undetermined error
Print failed; Error transferring image to printer
Print failed; Printer ioctl failed
Print failed; Unknown printer response
Summary of Contents for ESA Series
Page 6: ...6 Contents Table of Contents ...
Page 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Page 33: ...33 Instrument Messages 1 Instrument Messages ...
Page 111: ...111 Functional Tests 2 Functional Tests ...
Page 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Page 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Page 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...